B-13
B.10 CO
2
Module Specifications
B.10.1 Sidestream CO
2
Module
CO
2
module
Measurement range and
accuracy
Measurement range Accuracy
0 to 40 mmHg ±2 mmHg
41 to 76 mmHg ±5 % of the actual reading
77 to 99 mmHg ±10 % of the actual reading
Measurement accuracy drift
According to the test method of the standard ISO
80601-2-55, the module meets the requirement for
measurement accuracy in this table.
Resolution 1 mmHg
Rise time
Adult water trap:
<400 ms@70 mL/min
<330 ms@100 mL/min
<300 ms@120 mL/min
<240 ms@150 mL/min
Neonatal water trap:
<400 ms@70 mL/min
<330 ms@100 mL/min
Total system response time
Using neonatal water trap, neonatal sampling line:
<7.5 s @ 100 mL/min
<8 s @ 70 mL/min
Using adult water trap, adult sampling line:
<7.5 s @ 150 mL/min
<8 s @ 120 mL/min
<8.5 s @ 100 mL/min
<9.5 s @ 70 mL/min
Pump rate
Adult: 70 mL/min, 100 mL/min, 120 mL/min and 150 mL/min
optional.
Pediatric: 70 mL/min and 100 mL/min optional.
The flow control accuracy is ±15 % of the set value or ±15
mL/min, whichever is greater.
Water trap cleaning time
Adult water trap:
≥24 h@150 mL/min
≥48 h@70 mL/min
Neonatal water trap:
≥24 h@100 mL/min
≥48 h @70 mL/min