E-7
E.6 Alarm
Alarm Factory default setting
Paw high alarm limit 50 cmH
2
O
MV high alarm limit Adult: 7.4 L/min; pediatric: 3.2 L/min
MV low alarm limit Adult: 2.9 L/min; pediatric: 1.3 L/min
TVe high alarm limit Adult: 980 mL; pediatric: 210 mL
TVe low alarm limit Adult: 245 mL; pediatric: 55 mL
FiO
2
high alarm limit 100 %
FiO
2
low alarm limit 21 %
EtCO
2
low alarm limit Adult: 15 mmHg; pediatric: 20 mmHg
EtCO
2
high alarm limit 50 mmHg
SpO
2
high alarm limit 100 %
SpO
2
low alarm limit 90 %
Desat alarm limit 80 %
PR high alarm limit Adult: 120; pediatric: 160
PR low alarm limit Adult: 50; pediatric: 75
f high alarm limit OFF
Tapnea 15 s
Alarm volume 5
E.7 History Data
Trend log Factory default setting
Graphic trend-display group All
Graphic trend-zoom 10 min
Tabular trend-display group All
Tabular trend-interval 1 min
Event logbook-filter All Events
E.8 Special Functions
Special function Factory default setting
Nebulizer time 30 min
P-V tool-Pstart 3 cmH
2
O
P-V tool-Pmax 15 cmH
2
O
P-V tool-Flow 6 L/min
P-V tool-Vlimit 770 mL
Recruitment Tool-
ā³
Pinsp 5 cmH
2
O