B-8
RCexp 0.0 to 10.0 0.1 s
TVe/IBW 0 to 50 0.1 mL/kg
Tinsp 0.00 to 60.00 0.01 s
I:E 100:1 to 1:150 0.1 /
Monitored parameters (O
2
Therapy)
Continuous Flow 0 to 100 1 L/min
O
2
Concentration 15 to 100 1 Vol.%
B.7 Ventilator Accuracy
Control accuracy
O
2
% ± (3 Vol.% +1% of setting)
TV ± (10 mL +10% of setting) (BTPS)
f ±1 bpm
fsimv ±1 bpm
Tinsp ±0.1 s or ±10 % of setting, whichever is greater
I: E
2: 1 to 1: 4: ±10 % of setting
Other range: ±15 % of setting
Tslope ± (0.2 s+20 % of setting)
PEEP ± (2.0 cmH
2
O + 5 % of setting)
△Pinsp ± (2.0 cmH
2
O + 5 % of setting)
△Psupp ± (2.0 cmH
2
O + 5 % of setting)
Phigh ± (2.0 cmH
2
O + 5 % of setting)
Plow ± (2.0 cmH
2
O + 5 % of setting)
Thigh ± 0.2 s or ±10 % of setting, whichever is greater
Tlow ± 0.2 s or ±10 % of setting, whichever is greater
Trigger ± (1.0 cmH
2
O + 10 % of setting)
± (1.0 L/min + 10 % of setting)
△int.PEEP ± (2.0 cmH
2
O + 5 % of setting)
Exp% ± 10 %
fapnea ± 1 bpm
△Papnea ± (2.0 cmH
2
O + 5 % of setting)
TVapnea ± (10 mL +10 % of setting) (BTPS)
Apnea Tinsp ± 0.1 s or ±10 % of setting, whichever is greater
Tpause(%) ± 0.1 s or ±10 % of setting, whichever is greater
MV% ±10% (absolute error) or ± 10% of set value, whichever is greater