6-4
6.3.2 Getting Patient Information from the ADT Server
The ventilator can connect with the Admit-Discharge-Transfer (ADT) server through the
eGateway, and the ventilator can load the patient information from ADT server.
To load patient information from the ADT server, perform the following procedure:
1. Connect the network cables.
2. Select [Setup]→[Maintain]→ [User] →Enter user password→[Interface Setting].
3. Select [LAN Setup] tab, set [IP Config. Method], [IP Address], [Subnet Mask] and
[Gateway] in the opened interface.
4. Select [eGateway] tab and set the [eGateway] to [ON] in the opened interface. Then set
the [IP] of eGateway and ADT. Normally, there is no need to set the [Port], but you can
change it as required.
5. Ensure the network status is [Connected] in the [eGateway] tab.
6. Enter Standby status.
7. Select
, input [Patient ID] and [Visit Number] in the opened interface.
8. Select [Query]. Then list pops up, including all the patients that meet the query criteria.
9. Select a patient from the patient list, and then select [Import]. The imported data
includes patient ID, visit number, first name, last name, bed number, Room number,
department, and facility.
NOTE
The IP address of the ventilator, eGateway and ADT must be on the same subnet.
When the [eGateway] is set to [ON], the ventilator can send the patient
information, ventilation mode, ventilation type, monitored paremeters, controlled
parameters, waveforms and alarm limits data to the eGateway.
6.4 Ventilation Type
The ventilator provides two ventilation types: invasive and non-invasive.
WARNING
Check the alarm limit settings after switching over from NIV to Invasive.