Safety Precautions
9
5. Notes on using the focus knobs
• Never rotate the left and right focus knobs on the microscopes in the opposite directions at the same
time. Doing so may damage the product.
• Do not rotate the focus knobs past their limit. Doing so may damage the product.
6. Protecting the ports.
The microscope has multiple ports. To keep out extraneous light and dust, attach the provided caps to
the ports that are not in use.
7. Caution on motorized devices
Do not force movements on the motorized parts by hand.
8. Note on vibrations caused by motorized operations
This product is designed to minimize the amount of vibration during motorized operation. However,
depending on the application, the vibration may have an effect on the microscopy.
9. Note on the Z-axis position display
Ti-E and Ti-E/B microscopes can display the Z-axis position on the front status display panel. Note,
however, that the precision of the value is not guaranteed, as Ti-E and Ti-E/B are not designed as
measurement machines.