EasyManuals Logo
Home>Rigol>Test Equipment>DS2302A

Rigol DS2302A User Manual

Rigol DS2302A
335 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #294 background imageLoading...
Page #294 background image
RIGOL Chapter 15 System Function Setting
MSO2000A/DS2000A User’s Guide
15-12
Self-calibration
The self-calibration program can quickly make the oscilloscope reach the best
working state to get the most precise measurement values. You can perform
self-calibration at any time especially when the change of the environment
temperature is up to or more than 5 ℃. Make sure that the oscilloscope has been
warmed up or operated for more than 30 minutes before the self-calibration.
Disconnect all the input channel connections, and then press Utility  Self-Cal to
open the self-calibration menu and self-calibration interface.
Press Start and the oscilloscope will start to execute the self-calibration program.
Press Exit to give up the self-calibration operation at any time and return to the
previous menu.
Note: Most of the keys are disabled during the self-calibration.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

Table of Contents

Other manuals for Rigol DS2302A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS2302A and is the answer not in the manual?

Rigol DS2302A Specifications

General IconGeneral
BrandRigol
ModelDS2302A
CategoryTest Equipment
LanguageEnglish

Related product manuals