Commissioning
7.4 5-minute quick start with SIWATOOL
SIWAREX WP321
Manual, 07/2014, A5E33715669A-02
39
5-minute quick start with SIWATOOL
General information on using the SIWATOOL program can be found in chapter "Service with
the SIWATOOL program (Page 42)".
For the quick setup, the required parameters are marked in bold font in data records DR 3
and DR 10. The procedure is described below.
Figure 7-8 Quick setup with marked parameters
Service mode must be activated in order to change the calibration parameters. You can find
the command in the "Service Commands" group (spanner icon).
The quick setup is based on the standard settings of the weighing module. The factory
setting must therefore be reset prior to the quick start. Firstly, service mode is activated; the
factory setting is subsequently loaded using the "Load factory setting (11)" command.