Scale parameters and functions
All parameters are set to default values in the factory. You can restore the configuration to
factory settings using the "Load factory settings" command.
The scales are ready for operation with the default parameter settings. You do not need to
re-enter all parameters. The advantage of this solution is that you can decide the default
values to be retain and the parameters that need to be adapted for your application.
All parameters are divided into data records (DR). The data records are organized in steps
(tasks) to be implemented during commissioning or during the process.
The scale functions governed by the parameters are also described in the parameter
description below.
First, the parameters of a given data record are displayed in a table. The detailed parameter
description for the parameters of this data record then follows.
When it receives new parameters, the SIWAREX module runs a validation check. In the
event of a parameter assignment error, the data record is not applied (not saved) by the
SIWAREX module and a data/operator error is reported.
DR 2 is a special data record used to transfer commands to the SIWAREX module by
SIWATOOL.