Messages
9.5 Message list
SIWAREX WP321
88 Manual, 07/2014, A5E33715669A-02
Data and operating errors message list
Data and operating errors
(code 5000 to 8999)
5000 Data and operating error exists 5000 Group error, a bit is set in the data and operating error bits
5001 Command code or data record un-
known
5001 Command code or data record is not known with current
application
5003 Cannot exit service mode 5003 Cannot exit service mode; calibration incomplete
5004 Command or data transmission only
available in service mode
5004 Activation o
f service mode is required to execute command or
transmit data
5005 Command or data transmission not
possible because service mode is active
5005 Command can currently not be executed because service
mode is active
5006 Command or data transmission not
possible because of BUSY
5006 Command can currently not be executed because module is
BUSY (data record or command transmission already active,
…)
5007 Command or data transmission not
possible because module is faulty or
SIMATIC CPU stop
5007 Command can currently not be executed because of a prob-
lem or SIMATIC CPU stop
5101 Command is not permissible in this
operating state
5101 Command is not permissible in the current operating state
5102 Command not possible because
there is no standstill
5102 Scale command (set to zero, tare, log, … ) cannot be execut-
ed because standstill is missing.
5104 Command not possible because
range is exceeded
5104 Command (e.g. set to zero, tare, calibrate command) cannot
be executed because the permitted range has been exceed-
ed.
5105 Load cell parameters not plausible 5105 Load cell parameters in data record DR 10(a) or data record
DR 11(d) not plausible (number, supporting points, load spec-
ifications, etc.).
5107 Characteristic cannot be moved 5107 Characteristic cannot be moved due to possibility of range
violation.
7000 Permitted number range violated 7000 The permitted number range, such as for weight values, was
violated.
7002 Specifications of string lengths not
plausible
7002 The string header in a specified string variable is not plausi-
ble.
7007 The calibration weights or calibration
digits are not plausible
7007 Specifications for calibration weights or digits in data record
DR 3 are incorrect (minimum distance, reversal of incline).
7008 Zeroing or tare parameter not plausi-
ble
7008 The specifications (data record DR 3) or tare specifications
(data record DR 15) are not plausible.
7009 Standstill range / standstill wait time 7009 Standstill range or standstill wait time are not plausible.
7010 Scale interval / rounding 7010 Scale interval or selection for rounding to decimal places is
not plausible.
7011 Filter parameters 7011 Specification of filter parameters is not plausible.