Scale parameters and functions
8.3 DR 3 calibration parameters
SIWAREX WP321
50 Manual, 07/2014, A5E33715669A-02
Maximum negative zeroing limit
Zeroing defines the current weight of the scales as zero.
You can limit the effect of the zeroing function. The limitation is based not on the current
gross weight, but rather on the weight which the scales would display had there been no
zeroing (time of scale calibration).
Maximum positive zeroing limit
You can limit the effect of the zeroing function. The limitation is based not on the current
weight, but rather on the weight which the scales would display had there been no zeroing
(time of scale calibration).