EasyManuals Logo
Home>Tektronix>Test Equipment>434

Tektronix 434 User Manual

Tektronix 434
222 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #102 background imageLoading...
Page #102 background image
Calibration434
HO RIZONTAL SYSTEM C ALIBRATION
Equipment Required
1. Test Oscilloscope
4. 42-Inch 50 £2 BNC Cable
2. Time-Mark Generator
5. 3-Inch Screwdriver
3. Standard Amplitude Calibrator
6. Low-Capacitance Screwdriver
Control Settings
Preset instrument controls to the settings given under
Preliminary Control Settings except as follows:
Sweep MODE AUTO
INTENSITY Visible Display
34. Adjust Sweep Generator Offset Adjustment
a. Connect a IX probe to the test oscilloscope vertical
input.
b. Connect the tip of the IX probe to TP701 (see Fig.
5-12).
c. Set the test oscilloscope input coupling to ground and
establish a 0 V DC reference level.
d. Set the test oscilloscope input coupling to DC and
the Volts/D iv to 5 mV.
e. CHECKTest oscilloscope display is w ithin two divi
sions of the 0 V DC reference level.
f. ADJUSTSweep Generator Offset adjustment R703
(see Fig. 5-12) to center the test oscilloscope display
around the 0 V DC reference level.
g. Disconnect all test equipment.
35. Adjust Sweep Cal Adjustment
a. Connect a Time-Mark Generator to the CH 1 input
connector via a 42-inch 50 £2 BNC cable.
M agnifier
Registration
R863 H
W k H orizontal
R817
Test P oint
TP701
ac-iSRzr
# isr?0'' T fik r iU 9 : -
.
Sweep G enerator
O ffse t R703
5-24
Fig. 5-12. Location o f ho rizontal system adjustments and test po ints.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix 434 and is the answer not in the manual?

Tektronix 434 Specifications

General IconGeneral
BrandTektronix
Model434
CategoryTest Equipment
LanguageEnglish

Related product manuals