EasyManuals Logo

Tektronix 434 User Manual

Tektronix 434
222 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #94 background imageLoading...
Page #94 background image
Calibration434
21. Check CH 2 Variable VO LTS/DIV Range and
Attenuator Accuracy
a. Remove the Standard A m plitude Calibrator signal
from the CH 1 input connector and connect it to the CH 2
input connector and set the Vertical Mode to CH 2.
b. W ith the CH 2 VO LTS/DIV switch set to 10 mV, set
the Standard A m plitude Calibrator fo r a 50 m illivolt output
signal amplitude.
c. W ith an exactly 5-division amplitude display, rotate
the CH 2 Variable V O LTS/D IV control.
d. CHECKCRT display can be reduced in amplitude to
two divisions or less.
e. Set the CH 2 Variable VO LTS /D IV control to CAL.
f. CHECKUsing the CH 2 VO LTS /DIV switch and the
Standard Am plitude Calibrator settings given in Table 5-1,
check to see if the vertical deflection factor accuracy for
each position of the CH 2 VO LTS/DIV switch is w ithin 3%.
22. Check ADD Mode Operation
a. A pply the Standard Am plitude Calibrator output
signal to both the CH 1 and CH 2 input connectors via a
BNC T and tw o 42-inch 50 SI BNC cables.
b. Set both VO LTS/DIV switches to .1 V and the Ver
tical Mode to ADD.
c. Set the Standard Am plitude Calibrator fo r a .2 volt
output signal.
d. CHECKCRT display fo r 4 divisions, ±0.12 division
of deflection.
e. Disconnect all test equipment.
23. Adjust Channel 1 VO LTS/DIV Switch Com
pensation
a. Set the Channel 1 V O LTS /D IV switch to 10 mV, the
Vertical Mode to CH 1, and the CH 1 Input Coupling to
DC.
b. Connect the high-amplitude output o f the Square-
Wave Generator (Type 106) to the CH 1 input connector
via a GR to BNC adapter, a 42-inch 50 SI BNC cable, a 50
SI BNC 10X attenuator, a 50 SI BNC termination, and a 24
picofarad input normalizer.
c. Adjust the output of the Square-Wave Generator fo r a
display five divisions in amplitude and a repetition rate of
one kilohertz.
d. Adjust the TRIGGER LEVEL control fo r a stable
display.
e. CH E C K-C R T display at each Channel 1 VO LTS/DIV
switch position for optimum square corner and fla t top.
Typically less than +2%, 2%, or a total of 2% peak to peak
aberration (+3%, 3%, or total of 3% peak to peak if the
measurement is being made outside of the +20°C to +30°C
temperature range). Remove the 10X attenuator and the 50
£2 BNC term ination as necessary to maintain a five-division
display amplitude.
f. ADJU STChannel 1 VO LTS/D IV switch compensa
tion as given in Table 5-2. Adjust fo r optimum square
corner and flat top. Remove the 10X attenuator and 50 £2
BNC term ination as necessary to maintain a five-division
display amplitude. Fig. 5-8 shows the location of the
variable compensation adjustments.
TABLE 5-2
VO LTS/D IV Switch Compensation
VO LTS /D IV
Switch Setting
Attenuator
Adjusted fo r optimum
Compensated Square
Corner
Flat
Top
10 mV ±1
C9 for Channel 1
C19 for Channel 2
5 mV Check
If error is greater than 2%,
com prom ise setting at 1
mV, 2 mV, 5 m V, and 10
m V fo r best o v e ra ll
response.
2 mV Check
1 mV
Check
20 mV
-±2 C l 06 | C l07
50 mV ±5
C110 j C111
.1 v
±10
Cl 14 C l 15
Remove external 10X attenuator from generator
.2 V
Check
If error is greater than 2%,
compromise setting at .1 V,
.2 V, and .5 V fo r best over
all response.
.5 V
Check
Remove 50 SI termination
1 V
±100 C l 18 Cl 19
2 V Check
If error is greater than 2%,
5 V
Check
compromise setting at 1 V,
2 V, 5 V, and 10 V for best
overall response.
10 V
Check
5-16

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix 434 and is the answer not in the manual?

Tektronix 434 Specifications

General IconGeneral
BrandTektronix
Model434
CategoryTest Equipment
LanguageEnglish

Related product manuals