VOLTAG E AND W AVEFORM TEST CONDITIONS
434
Typical voltage measurements and waveform photographs
were obtained under the following conditions unless noted
otherwise on the individual diagrams. Voltage measure
ments were taken with no signal applied to the vertical
input. Waveform photographs were taken with the PROBE
CAL 0.6 V 1 kHz signal applied to the CH 1 input
connector.
Variable
Input Coupling
Vertical Mode
Trigger Controls
Test Oscilloscope
SLOPE
Frequency response DC to 40 MHz.
LEVEL
Deflection factor
5 millivolts to 50 volts/division.
SOURCE
Probe ground 434 chassis ground.
EXT ATTEN
Trigger source
External to indicate true time rela
tionship between signals.
COUPLING
Recommended type
Tektronix 7504 with 7A16 plug-in
unit and P6052 Probe.
Sweep Controls
MODE
Voltm eter
TIME/DIV
Type Digital multimeter.
MAG
Input impedance
10 megohms.
Variable
CAL
DC
CH 1
Midrange
CH 1
1:1
DC
AUTO
1 ms
XI
Midrange
Range 0 to 1 kilovolt.
Reference voltage 434 chassis ground.
Recommended type Tektronix 7504 with 7D13 Digital
(as used for voltages Multimeter Plug-In Unit,
on diagrams)
Storage Controls (upper and lower if applicable)
STORE Non-store (button out)
ENHANCE Off (button out)
ENHANCE LEVEL Fully counterclockwise
434 Conditions
POWER/ Pulled out (on) and set for normal
INTENSITY display intensity.
Vertical Controls (both channels if applicable)
POSITION Midrange
OUT: 5 MHz BW
Pushed in
OUT: INVERT
Push in
VOLTS/DIV
.1 V
82
®