EasyManuals Logo
Home>Tektronix>Test Equipment>434

Tektronix 434 User Manual

Tektronix 434
222 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #93 background imageLoading...
Page #93 background image
Calibration434
e. Set the CH 2 VO LTS/D IV switch to 10 mV.
f. Rotate the CH 2 Variable V O LTS/D IV control.
g. CHECKCRT display for not more than 0.2 division
of trace shift when rotating the CH 2 Variable V O LTS /D IV
control.
h. ADJU STVariable Balance adjust R312 (see Fig.
5-7) for minimum trace shift when rotating the CH 2
Variable V O LTS /D IV control.
i. Set the CH 2 VO LTS /DIV switch to 10 m V and the
CH 2 Variable VO LTS/D IV control to CAL.
j. Position the free-running trace to the center hori
zontal graticule line.
k. Set the INVER T switch to the INVERT position
(button out).
I. CHECKCRT display for not more than 0.2 division
of trace shift when inverting the Channel 2 display.
m. ADJUSTInvert Balance adjustment R348 (see Fig.
5-7) for m inimum trace shift when inverting the Channel 2
display.
n. Push the INVERT button in.
19. Adjust CH 1 and CH 2 GAIN
a. Set input coupling for both channels to DC.
b. Connect the output of the Standard Am plitude Cali
brator to the CH 2 input connector via a 42-inch BNC
cable.
c. Set the Standard Am plitude Calibrator for a 50 m illi
volt signal output amplitude.
d. CHECKCRT display for 5 divisions, ±0.15 division
of deflection.
e. ADJUSTCH 2 GAIN adjustment (located on fro n t
panel) for exactly 5 divisions of deflection.
f. Change Vertical Mode to CH 1.
g. Remove the Standard Am plitude Calibrator signal
from the CH 2 input connector and connect it to the CH 1
input connector.
h. CHECKCRT display for 5 divisions, ±0.15 division
of deflection.
i. ADJUSTCH 1 GAIN adjustment (located on fro nt
panel) for exactly 5 divisions of deflection.
20. Check CH 1 Variable VO LTS/D IV Range and
Attenuator Accuracy
a. With exactly a 5-division amplitude display, rotate
the CH 1 Variable V O LTS/D IV control.
b. CHECKCRT display can be reduced in amplitude to
two divisions or less.
c. Set the CH 1 Variable VO LTS/D IV control to CAL.
d. CHECKUsing the CH 1 V O LTS/DIV switch and the
Standard Am plitude Calibrator settings given in Table 5-1,
check to see if the vertical deflection factor accuracy for
each position o f the CH 1 VO LTS/DIV switch is w ithin 3%.
TABLE 5-1
Vertical Deflection Accuracy
VO LTS /D IV
Switch
Setting
Standard
Am plitude
Calibrator
Output
Vertical
Deflection
in
Divisions
Maximum
Error
For +3%
Accuracy
1 mV
5 m illivolts
5
±0.15 division
2 mV
10 m illivolts
5
±0.15 division
5 mV 20 m illivolts
4
±0.12 division
10 mV 50 m illivolts 5
Previously set
in step 19.
20 mV 0.1 volt
5
±0.15 division
50 mV 0.2 volt
4 ±0.12 division
.1 V
0.5 volt 5
±0.15 division
.2 V 1 volt 5 ±0.15 division
.5 V 2 volts
4 ±0.12 division
1 V
5 volts 5 ±0.15 division
2 V 10 volts 5 ±0.15 division
5 V 20 volts 4
±0.12 division
10 V 50 volts 5
±0.15 division
5-15

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix 434 and is the answer not in the manual?

Tektronix 434 Specifications

General IconGeneral
BrandTektronix
Model434
CategoryTest Equipment
LanguageEnglish

Related product manuals