Syntax and Commands
3-32 AFG3000 Series Arbitrary/Function Generators Reference Manual
OUTPut[1|2]:IMPedance
The OUTPut:IMPedance command sets the output load impedance for the specified
channel. The specified value is used for amplitude, offset, and high/low level
settings. You can set the impedance to any value from 1 to 10 k. The default
value is 50 .
The OUTPut:IMPedance? command returns the current load impedance setting in
ohms. If the load impedance is set to INFinity, the query command returns
“9.9E+37”.
Group
Output
Syntax
OUTPut[1|2]:IMPedance{<ohms>|INFinity|MINimum|MAXimum}
OUTPut[1|2]:IMPedance?[MINimum|MAXimum]
Arguments
<ohms>::=<NR3>[<units>]
where
<units>::=OHM
INFinity sets the load impedance to >10 kΩ.
MINimum sets the load impedance to 1 .
MAXimum sets the load impedance to 10 k.
Returns
<ohms>::=<NR3>
Examples
OUTPut1:IMPedance MAXimum
sets the CH 1 load impedance to 10 k .

Table of Contents

Other manuals for Tektronix AFG3000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix AFG3000 Series and is the answer not in the manual?

Tektronix AFG3000 Series Specifications

General IconGeneral
BrandTektronix
ModelAFG3000 Series
CategoryPortable Generator
LanguageEnglish

Related product manuals