Syntax and Commands
AFG3000 Series Arbitrary/Function Generators Reference Manual 3-35
*PSC
This command sets and queries the power-on status flag that controls the automatic
power-on execution of SRER and ESER. When *PSC is true, SRER and ESER are
set to 0 at power-on. When *PSC is false, the current values in the SRER and ESER
are preserved in nonvolatile memory when power is shut off and are restored at
power-on.
Group
Status
Syntax
*PSC <NR1>
*PSC?
Arguments
<NR1>=0 sets the power-on status clear flag to false, disables the power-on clear,
and allows the instrument to possibly assert SRQ after power-on.
<NR1>0 sets the power-on status clear flag true. Sending *PSC 1 therefore
enables the power-on status clear and prevents any SRQ assertion after power-on.
Returns
<NR1>
Examples
*PSC 0
sets the power-on status clear flag to false.
*RCL (No Query Form)
This command restores the state of the instrument from a copy of the settings stored
in the setup memory. The settings are stored using the *SAV command. If the spec-
ified setup memory is deleted, this command causes an error.
Group
Memory
Related Commands
*SAV
Syntax
*RCL {0|1|2|3|4}
Arguments
0, 1, 2, 3, or 4 specifies the location of setup memory.
Examples
*RCL 3
restores the instrument from a copy of the settings stored in memory location 3.