Syntax and Commands
3-76 AFG3000 Series Arbitrary/Function Generators Reference Manual
[SOURce[1|2]]:PWM[:DEViation]:DCYCle
This command sets or queries the PWM deviation in percent for the specified
channel.
The setting range must meet the following conditions:
Deviation ≤ Pulse Width – PWmin
Deviation ≤ Pulse Period – Pulse Width – PWmin
Deviation ≤ Pulse Width – 0.8 × (Leading Edge Time + Trailing Edge Time)
Deviation ≤ Pulse Period – Pulse Width – 0.8 × (Leading Edge Time + Trailing
Edge Time)
where PWmin is the minimum pulse width.
Group
Source
Syntax
[SOURce[1|2]]:PWM[:DEViation]:DCYCle {<percent>|MINimum|MAXimum}
[SOURce[1|2]]:PWM[:DEViation]:DCYCle? {MINimum|MAXimum}
Arguments
<percent>::=<NR2>[<units>]
where
<NR2> is the PWM deviation.
<units>::=PCT
Returns
<percent>
Examples
SOURce1:PWM[:DEViation]:DCYCle 5.0
sets the CH 1 PWM deviation to 5.0%.