Acquiring Waveforms
3-50
CSA7000B Series & TDS7000B Series Instruments User Manual
H FastFrame reduces the time required before the trigger is rearmed, while
preserving the individual subrecords, a detail lost in Fast Acquisitions, which
writes all acquired records to a single pixel map.
Use the procedure that follows to set up FastFrame mode acquisitions.
Overview To set FastFrame mode Control elements and resources
Prerequisites
1. The horizontal and vertical controls must be set up.
Triggering should also be set up.
See page 3--29 for acquisition setup and
page 3--63 for trigger setup.
To set
FastFrame
mode
2. Touch the Horiz button. Select the Acquisition tab from
the Horiz/Acq control window. Touch FastFrame Setup
to display the FastFrame Setup control window.
3. Touch FastFrame to toggle FastFrame to On.
Set frame
length
4. Touch Rec Length, and set the number of samples per
frame.
Record length is the number of samples in each
acquisition.
To Set FastFrame Mode

Table of Contents

Other manuals for Tektronix CSA7404B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix CSA7404B and is the answer not in the manual?

Tektronix CSA7404B Specifications

General IconGeneral
BrandTektronix
ModelCSA7404B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals