Measuring Waveforms
3- 152
CSA7000B Series & TDS7000B Series Instruments User Manual
Cursors can measure channel, reference, and math waveforms. You must set the
source of each cursor explicitly in the Cursor Setup control window.
Cursors are not available with histograms, XY, or XYZ modes.
Cursor operation is easy, you move the cursors on screen and read the results in the
cursor readouts. The following key points will help you use the cursors effectively:
Cursor Types. The cursor types are described in Table 3--9 on page 3--151. There
are two cursors displayed for all types, Cursor 1 and Cursor 2. You can move
cursors with the multipurpose knobs or the cursor position controls in the Cursor
Setup control window.
+ 2 divisions at 20 mv/div.
Figure 3- 38: Horizontal cursors measure amplitudes
Cursors are Display-Limited. You cannot move a cursor off screen. Also, if you
resize waveforms, the cursors do not track. That is, a cursor stays at its screen
position, ignoring changes to horizontal and vertical scale and position, and
vertical offset (waveform cursors will track a waveform vertically).
Cursors Ignore the Selected Waveform. Each cursor measures its source, defined
in the Cursors Setup dialog box. Selecting a waveform for scaling on screen (by
pushing the CH 3 front-panel button, for example) does not change the source
that each cursor measures.
Using Cursors