Appendix A: Specifications
A-28
CSA7000B Series & TDS7000B Series Instruments User Manual
Table A- 13: Certifications and compliances
Category Standards or description
EC Declaration of Conformity --
EMC
Meets intent of Directive 89/336/EEC for Electromagnetic Compatibility. Compliance was
demonstrated to the following specifications as listed in the Official Journal of the European Union:
EN 61326 Emissions
1, 3, 4
Class A Radiated and Conducted Emissions
EN 61326 Immunity
1, 4
IEC 1000-4-2 Electrostatic Discharge Immunity
±4 kV contact discharge, ±4 kV air discharge
IEC 1000-4-3 RF field immunity
3V/m,80MHzto1GHz,
80% amplitude modulated with a 1 kHz sinewave
2
IEC 1000-4-4 Electrical Fast Transient/Burst Immunity
1 kV on AC mains, 500 V on I/O cables
IEC 1000-4-5 AC Surge Immunity
1 kV differential mode, 2 kV common mode
IEC 1000-4-6 RF Conducted Immunity
3 V, 150 kHz to 80 MHz, 80% amplitude modulated with a
1 kHz sinewave
2
IEC 1000-4-11 AC Mains Voltage Dips and Interruption Immunity
100% reduction for one cycle
EN 61000-3-2 Power Harmonic Current Emissions
EN 61000-3-3 Voltage Changes, Fluctuations, and Flicker
1
If interconnect cables are used, they must be low-EMI shielded cables such as the following
Tekt roni x part numbers or their equivalents: 012-0991-00, 012-0991-01, 012-0991-02 or
012-0991-03 GPIB Cable; 012-1213-00 (or CA part number 0294-9) RS-232 Cable;
012-1214-00 Centronics Cable; or LCOM part number CTL3VGAMM-5 VGA Cable.
2
Under these conditions, the specifications are amended as follows:
10 mV/division to 1 V/division: ≤0.4 division waveform displacement or ≤0.8 division
increase in peak-to-peak noise
5 mV/division and 2 mV/division, typical: ≤8 mV increase in peak-to-peak noise
3
Radiated emissions may exceed the levels specified in EN 61326 when this instrument is
connected to a test object.
4
Test ed in accordance with EN 61326 Annex D.
Australia/New Zealand Declara-
tion of Conformity — EMC
Complies with EMC provision of Radiocommunications Act per the following standard(s):
AS/NZS 2064.1/2 Industrial, Scientific, and Medical Equipment: 1992, Class A

Table of Contents

Other manuals for Tektronix CSA7404B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix CSA7404B and is the answer not in the manual?

Tektronix CSA7404B Specifications

General IconGeneral
BrandTektronix
ModelCSA7404B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals