CSA7000B Series & TDS7000B Series Instruments User Manual
A-1
Appendix A: Specifications
This chapter contains the specifications for the CSA7000B Series Communica-
tions Signal Analyzers and the TDS7000B Series Digital Phosphor Oscillo-
scopes. All specifications are guaranteed unless labeled “typical.” Typical
specifications are provided for your convenience but are not guaranteed.
Specifications that are marked with the n symbol are checked in chapter four ,
Performance Verification, of the service manual.
All specifications apply to all models unless noted otherwise. To meet specifica-
tions, the following conditions must be met:
H The instrument must have been calibrated in an ambient temperature
between 20 °C and 30 °C(68°F and 86 °F).
H The instrument must be operating within the environmental limits listed in
Table A--12 on page A--27.
H The instrument must be powered from a source that meets the specifications
listedinTableA--10onpageA--25.
H The instrument must have been operating continuously for at least 20 min-
utes within the specified operating temperature range.
H You must perform the Signal P ath Compensation procedure after the
20-minute warm-up period, and if the ambient temperature changes more
than 5 °C(41°F), repeat the procedure. See Optimizing Measurement
Accuracy on page 3--160 for instructions to perform this procedure.