Triggering
3- 108
CSA7000B Series & TDS7000B Series Instruments User Manual
The flow diagram in Figure 3--31 summarizes all combinations of triggering and
horizontal delay.
Wait for
A(Main)
trigger
B trigger after delay with
horizontal delay off
Wait for B
trigger
event
1
Acquire
posttrigger
data
Wait the
user-specified
number of B trigger
events
1
Wait user-specified
trigger delay time
Wait
user-specified
horizontal
delay time
Wait the
user-specified
number of B trigger
events
1
B trigger on events with
horizontal delay off
B trigger on events with
horizontal delay on
Wait
user-specified
horizontal
delay time
B trigger after delay with
horizontal delay on
Wait for B
trigger
event
1
Wait
user-specified
horizontal
delay time
Trigger on A Only with
horizontal delay on
Trigger on A Only with
horizontal delay off
Reset
1
1
If Reset is set, waiting on B event(s) only occurs until Reset condition you specify is satisfied. At that point, the Trigger system resets, and the sequence
restarts with waiting for the A Event.
Figure 3- 31: Trigger and Horizontal Delay summar y