Triggering
CSA7000B Series & TDS7000B Series Instruments User Manual
3- 109
Sequential Trigger Reset. You can choose three methods to reset sequential
triggers:
H None does not reset sequential triggers. Triggers function as previously
explained
H Timeout resets sequential triggers if the time you set expires before the B
trigger event occurs
H Transition resets sequential triggers if a signal transition you select occurs
before the B trigger event occurs
H State resets sequential triggers if a signal state you select occurs before the B
trigger event occurs
Use the procedure that follows when setting up the instrument to trigger on a
sequence. For more information, display online help while performing the
procedure.
Overview To trigger on a sequence Control elements and resources
Prerequisites 1. The inst rum ent must be installed with a signal
connected to an input channel. Acquisition system
should be set to Run, and the vertical and horizontal
controls should be set appropriately for the signal to be
acquired.
See page 3--23 for acquisition setup
To trigger on
a (main) only
1. From the toolbar, touch Trig and select the A-->B Seq
tab of the Trigger control window.
2. Touch A Only to turn off sequential triggering.
To Trigger on a Sequence

Table of Contents

Other manuals for Tektronix CSA7404B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix CSA7404B and is the answer not in the manual?

Tektronix CSA7404B Specifications

General IconGeneral
BrandTektronix
ModelCSA7404B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals