Triggering
3-110
CSA7000B Series & TDS7000B Series Instruments User Manual
Overview Control elements and resourcesTo trigger on a sequence (Cont.)
To trigger on
B after time
1. To set the time base to run after an A trigger, a trigger
delay, and a B trigger, from the toolbar, touch Trig, and
select the A-->B Seq tab of the Trigger control window
2. TouchTrigAfterTime.
3. To set the trigger delay, touch Trig Delay, and use the
multipurpose knob or keypad to set the time.
4. To set the B trigger level, touch B Trig Level, and use
the multipurpose knob or keypad to set the level.
5. If using any other trigger type, see To set up B triggering
on page 3--115.
To trigger on
B events
1. To set the time base to trigger after an A trigger and a
specified number of B trigger events, from the toolbar,
touch Trig, and select the A-->B Seq tab of the Trigger
control window.
2. Touch A Then B Trig on nth Event.
3. To set the number of B trigger events, touch Trig Event,
and use the multipurpose knob, keypad, or up and down
arrows to set the number of events.
4. To set the B trigger level, touch B Trig Level, and use
the multipurpose knob or keypad to set the level.
5. If using any other trigger type, see To set up B triggering
on page 3--115.