Triggering
3-112
CSA7000B Series & TDS7000B Series Instruments User Manual
Overview Control elements and resourcesTo trigger on a sequence (Cont.)
Reset timeout
B event
End of trigger delay
A event
To reset
sequential
triggers (Cont.)
3. If the Transition Type of Trigger Reset is selected, touch
Source and select the source of the transition from the
list.
Touch Slope and select the slope of the transition from
the list.
Touch Level and use the multipurpose knobs or keypad
to set the level that the source must pass through to
reset the triggers.
In the example, the runt trigger A event occurs on the
falling edge of the pulse on Ch1. The trigger delay
expires 1 s after the falling edge of the pulse on Ch1.
Ch2 provides the B events, and Ch3 provides the reset
event. The reset event window is from the A event to the
reset event at the rising edge of Ch3.
If the trigger delay is increased from 1 sto2s, the
instrument will not trigger because the Ch3 reset event
would occur before the next Ch2 event.