Creating and Using Math Waveforms
CSA7000B Series & TDS7000B Series Instruments User Manual
3- 219
Overview Control elements and resourcesSpectral math example (cont.)
Measure the
test results
14. Use the multipurpose knobs or keypad to set the Curs1
Pos to 3.0 kHz and the Curs2 Pos to 11.0 kHz.
In this example, the cursors are now on the third and
eleventh harmonic of the probe compensation signal.
Read the frequencies from the cursor readouts.
15. Touch the Cursor Type Waveform button. Touch the
Cursor 2 button and then the Math 1 button.
Now in addition to the frequency at the cursor locations,
the cursor readout displays the amplitude at the cursor
locations. The readout also displays the difference in
frequency and amplitude between the cursor locations.
For more
information
16. For additional information on setting up and using
spectral math, see Defining Spectral Math Waveforms
starti ng on page 3--182.