MENU ITEM
Monitor/
Controller
The Display Meas item selects which measurements will
be presented on the display during normal operating
mode. Use the Units item to choose the units of measure:
ROOM STATUS displays the room mode as negative,
positive or no isolation.
ROOM PRESSURE displays the room mode and the
current measurement of room pressure differential.
ALL displays the room mode and all currently connected
measurements.
Room Status, Room
Pressure, All
NOTE: Measurements will still enable alarms if not on the
display. The measurement will not appear on the
display even when in alarm status if not so enabled.
The Display Avg item selects the display’s running
average period. The display-averaging period is the length
of time the face velocity has been averaged before being
displayed. The Display Avg item value may be set
between 0.5 and 40 seconds. The higher the averaging
value, the more stable the display.
1, 2, 3, 5, 10, 20, or
40 seconds
The Units item selects the unit of measure that the
controller displays all values (except calibration span).
These units display for all menu items setpoints, alarms,
flows, etc.
in. W.C., cfm, F
Pa, l/s, C
Pa, cmh, C
The Input 1 item selects the desired input type for Input1,
the room pressure sensor for Room 1
TSI Sensor, Pressure
Transducer
Go to the Input 1 menu to adjust parameters such as