EasyManuals Logo
Home>YOKOGAWA>Recording Equipment>Daqstation DX2000

YOKOGAWA Daqstation DX2000 User Manual

YOKOGAWA Daqstation DX2000
366 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #67 background imageLoading...
Page #67 background image
1-55
IM 04L42B01-01E
Overview of Functions
1
2
3
4
5
6
7
8
9
10
11
12
13
App
Index
StatusOutput
Outputs the status below with a relay contact signal (1 relay). You can select whether
each status is output to the relay. The relay is energized when the status occurs. You
cannot change this behavior.
For the setting procedure, see section 2.9.
Status Description CorrectiveAction
Status of the internal
memory or CF card
Error in the internal memory. Contact your nearest
YOKOGAWA dealer for
repairs.
WhentheautosavefunctiontotheCFcardisOn.
• ThefreespaceontheCFcard
dropped to 10% of the total size (only
when the media FIFO (see section 1.4)
is disabled).
Replace the CF card.
• ErrorintheCFcard. • ReplacetheCFcardwith
a normal one.
• FormattheCFcardonthe
DX (the data on the CF
card will be erased).
However, the status of the internal
memory is output when the CF card is
not inserted.
• 10MBorlessofavailablespace*
remaining in internal memory.
• Thenumberoffilesininternalmemory
for which Auto Save to the CF card
has not been completed has exceeded
390.
Insert a CF card.
WhentheautosavefunctiontotheCFcardisOff.
• 10MBorlessofavailablespace*
remaining in internal memory.
• Thenumberoffilesininternalmemory
for which Manual Save has not been
completed has exceeded 390.
Save the data in the internal
memory to the CF card.
Measurement error Error in the A/D converter. Contact your nearest
YOKOGAWA dealer for
repairs.
Burnout is detected. Replace the thermocouple
that has burned out.
Communication error A Modbus master or Modbus client
communication error occurred.
Check the error in the
Modbus master or Modbus
client screen and carry out
the corrective action.
A PROFIBUS-DP communication error
occurred.
Contact your nearest
YOKOGAWA dealer for
repairs.
Memory stop When the memory sampling is stopped. Start the data acquisition.
Alarm occurrence
(Release number 3
or later)
An alarm has occurred. Check the alarm.
* The internal memory’s “available space” refers to the following quantities.
• Unusedregions.
• RegionsofdataforwhichAutoSaveorManualSave(seepage1-33)hasbeen
completed.
RelayOperation
NO C NC
During normal operation
NO C NC
When specified status occurs
NO C NC
When power is turned OFF
1.9FAIL/StatusOutputFunction(/F1and/F2Options)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA Daqstation DX2000 and is the answer not in the manual?

YOKOGAWA Daqstation DX2000 Specifications

General IconGeneral
BrandYOKOGAWA
ModelDaqstation DX2000
CategoryRecording Equipment
LanguageEnglish

Related product manuals