6-10
IM 04L42B01-01E
6.4 Starting/StoppingtheRecordingandSaving
theMeasuredData,SavingtheMeasuredData
throughKeyOperation
Start the recording and save the measured data to the external storage medium.
For a description of the function, see section 1.4.
Procedure
• StartingtheRecording(MemoryStart)
Press START. The internal memory icon in the status display section changes from
the stop icon to memory sampling icon.
• Whenrecordingdisplaydataoreventdatainfreemode,recordingstarts.
• Whenrecordingeventdataintriggermode,theDXentersthetrigger-wait
condition.
* On DXs with the /AS1 advanced security option, recording will not start, even if you press
START in setting mode.
* For the procedures when using the batch function, see section 6.3. If you are using the multi
batch function (/BT2 option), see the DX1000/DX1000N/DX2000 Multi Batch (/BT2) User’s
Manual, IM04L41B01-03E.
• ApplyingaTriggertoStarttheRecording
Carry out the procedure below when the DX is waiting for a trigger.
TriggerthroughKeyOperation
The procedure below can be carried out when recording event data in trigger mode
and the DX is configured so that the start trigger is applied through key operation.
1.
Press FUNC.
The
FUNC
key menu appears.
2.
Press the Trigger soft key.
The recording starts.
TriggerbyanEvent(Eventactionfunctionmustbeconfigured.See
section7.1.)
Recording starts when an event occurs.
• AutomaticallySavingMeasuredData
Automatic saving takes place when Auto save is set to On (see section 6.2 for details).
The save destination is the CF card.
Have the CF card inserted in the slot at all times. While the memory sampling is in
progress, the measured data recorded in the internal memory is automatically saved
to the CF card.
Action when Media FIFO is not enabled: If data storage to the storage medium is not
complete such as due to insufficient free space, the unsaved data is saved the next
time the data is automatically saved.
• SavingtheDisplayDataorEventDataduringMemorySamplingthrough
KeyOperation
The save destination is the CF card.
This operation can be carried out when recording display data or when recording
event data in Free mode. Unsaved measured data is saved to the CF card.
1.
In the operation mode, press FUNC.
The FUNC key menu appears.