COMPRESS
166 Chapter 6 Command Reference
Syntax CALSTR string[,security_code]
string This is the alpha/numeric message that will be appended to the calibration RAM.
The string parameter must be enclosed in single or double quotes. The maximum
string length is 75 characters (the quotes enclosing the string are not counted as
characters).
security_code When the calibration RAM is secured (SECURE command) you must include the
security_code in order to write a message to the calibration RAM. (You can always
read the string using the CALSTR? command regardless of the security mode).
Refer to the SECURE command for information on securing and unsecuring the
calibration RAM.
Remarks • Query Command. The CALSTR? query command returns the character string
from the multimeter's calibration RAM. This is shown in the second example
below.
• Related Commands: CAL, CALNUM?, SCAL, SECURE
Examples CALSTR
OUTPUT 722;"CALSTR 'CALIBRATED 04/02/1987'"
CALSTR?
10 DIM A$[80] !DIMENSION STRING VARIABLE
20 OUTPUT 722; "CALSTR?" !READ THE STRING
30 ENTER 722;A$ !ENTER STRING
40 PRINT A$ !PRINT STRING
50 END
COMPRESS
Compress Subprogram. Removes the ASCII text of a specified subprogram
previously stored in memory. This saves memory space but removes the
subprogram from continuous memory (the subprogram will be destroyed when
power is removed).
Syntax COMPRESS name
name Subprogram name. A subprogram name may contain up to 10 characters. The
name can be alpha. alphanumeric, or an integer in the range of 0 to 127. Refer to
the SUB command for details.
Power-on name = none.
Default name = none; parameter required.
Remarks • To avoid memory fragmentation, compress each subprogram before
downloading other subprograms.
• You cannot store the COMPRESS command as part of a subprogram.

Table of Contents

Other manuals for Agilent Technologies 3458A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 3458A and is the answer not in the manual?

Agilent Technologies 3458A Specifications

General IconGeneral
BrandAgilent Technologies
Model3458A
CategoryMultimeter
LanguageEnglish

Related product manuals