Chapter 6 Command Reference 233
SETACV
the factory with its security code set to 3458.
new_code
This is the new security code. The code is an integer from -2.1E9 to 2.1E9. If the
number specified is not an integer, the multimeter rounds it to an integer value.
acal_secure
Allows you to secure autocalibration. The choices are:
Power-on acal_secure = Previously specified value (ON is the factory setting).
Default acal_secure = OFF.
Remarks • Specifying 0 for the new_code disables the security feature making it no longer
necessary to enter the security code to perform a calibration or autocal.
• The front panel's Last Entry key will not display the codes used in a previously
executed SECURE command.
• Related Commands: ACAL, CAL, CALNUM?, CALSTR, SCAL
Examples Changing the Code
OUTPUT 722;"SECURE 3458,4448,0N" !CHANGE FACTORY SECURITY CODE TO 4448,
!ENABLE AUTOCAL SECURITY
Disabling Security
OUTPUT 722;"SECURE 3458,0" !DISABLES SECURITY FOR CALIBRATION AND
AUTOCAL
SETACV
Set ACV. Selects the RMS conversion technique to be used for AC or AC+DC
voltage measurements.
Syntax SETACV [type]
type
The type parameter is used to select the measurement method: analog, random
acal_secure
Parameter
Numeric
Query
Equiv. Description
OFF 0 Disables autocal security; no code required for
autocal
ON 1 Enables autocal security; the security code is
required to perform autocal (see ACAL for
example).

Table of Contents

Other manuals for Agilent Technologies 3458A

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 3458A and is the answer not in the manual?

Agilent Technologies 3458A Specifications

General IconGeneral
Model3458A
ManufacturerAgilent Technologies
CategoryMultimeter
Digits8.5
Sampling Rate100, 000 readings/second
InterfaceGPIB

Summary

Safety Symbols and Warnings

General Safety Precautions (WARNINGS)

Outlines essential safety precautions for operating, servicing, and repairing the product to prevent injury or damage.

Chapter 2 Getting Started

Operating from the Front Panel

Covers using front panel keys, making measurements, changing functions, and controlling display settings.

Operating from Remote

Explains how to control the multimeter remotely via GPIB, including address management and command sending.

Chapter 3 Configuring for Measurements

Configuring for DC or Resistance Measurements

Details how to configure the multimeter for DC voltage, DC current, and 2-wire or 4-wire resistance measurements.

Configuring for AC Measurements

Explains how to configure the multimeter for AC voltage, AC current, frequency, or period measurements.

Chapter 4 Making Measurements

Triggering Measurements

Explains the three-event triggering hierarchy (arm, trigger, sample) and various event choices.

Increasing the Reading Rate

Discusses the multimeter's high-speed mode and factors affecting reading rate and transfer speed.

Math Operations

Explains real-time and post-process math operations, enabling/disabling them, and math registers.

Chapter 5 Digitizing

Digitizing Methods

Details DCV, Direct-Sampling, and Sub-sampling methods, summarizing their characteristics and signal paths.

Level Triggering

Describes how to specify voltage and slope for sampling initiation, with examples for DCV and direct-sampling.

Direct-Sampling

Explains direct-sampling using track-and-hold, its bandwidth, and specifying ranges via max._input parameter.

Sub-Sampling

Covers sub-sampling fundamentals, advantages, and how to specify effective interval and number of samples.

Chapter 6 Command Reference

ACAL

Instructs the multimeter to perform self-calibrations (ALL, DCV, AC, OHMS) and discusses autocalibration security.

PRESET

Configures the multimeter to one of three predefined states: NORM, FAST, or DIG for different operation modes.

SUB

Stores a series of commands as a subprogram, assigning a name for later execution.

TARM

Defines the trigger arm event to enable the trigger event and can be used for multiple measurement cycles.

TEST

Causes the multimeter to perform a series of internal self-tests to check hardware and software integrity.

TRIG

Specifies the trigger event that initiates a measurement, working with TARM and NRDGS.

Chapter 7 BASIC Language for the 3458A

Subprograms

Explains how to store, execute, and manage BASIC language subprograms for system control and automation.

Appendix B GPIB Commands

Appendix C Procedure to Lock Out Front/Rear Terminals and Guard Terminal Switches

Procedure

Outlines the steps for installing the switch lockout kit, including covers and pushrod removal.

Appendix D Optimizing Throughout and Reading Rate

Maximizing the Testing Speed

Covers strategies like tailoring communication paths, program memory, and state storage for optimal testing speed.

DC Volts, DC Current and Resistance

Explains the measurement paths (DCV, track-and-hold) and trade-offs for DC measurements.

AC Volts and AC Current

Details the three ACV measurement techniques (Analog, Synchronous, Random) and their trade-offs.

Optimizing the Testing Process Through Task Allocation

Discusses allocating tasks between the DMM and computer using math functions, memory, and program structure.

Appendix E High Resolution Digitizing With the 3458A

Speed with Resolution

Details the multimeter's flexibility in speed and resolution for audio frequency bandwidth.

Avoiding Aliasing

Provides methods to avoid signal distortion caused by aliasing, ensuring accurate waveform representation.

Choice of Two Measurement Paths

Describes the standard DCV path and the track-and-hold path for digitizing and sampling.