Appendix B GPIB Commands 307
TRIGGER (GET)
TRIGGER (GET)
If triggering is armed (see TARM command), the TRIGGER command (Group
Execute Trigger) triggers the multimeter once, and then holds triggering.
Syntax TRIGGER 7
TRIGGER 722
Remarks • The TRIGGER command generates a single trigger just as if the TRIG SGL
command was executed. It will not, however, trigger the multimeter if triggering
is not armed (TARM command).
• If subprogram memory execution is suspended by the PAUSE command
(multimeter command set), the TRIGGER command resumes subprogram
execution but does not generate a single trigger.
Examples TRIGGER 7 !SENDS GROUP EXECUTE TRIGGER (GET)
TRIGGER 722 !SENDS GROUP EXECUTE TRIGGER (GET) TO THE DEVICE AT ADDRESS 22