70 Chapter 3 Configuring for Measurements
percent for the synchronous conversion method or 0.4 percent for the random
conversion method.) The following program selects AC voltage
measurements using the synchronous sampling conversion. The maximum
expected input voltage is 10 volts and a %_resolution parameter of .1 selects
5.5 digits resulting in an actual resolution of l mV.
10 OUTPUT 722; "SETACV SYNC"
20 OUTPUT 722;"ACV 10, .1"
30 END
Configuring for Ratio Measurements
For ratio measurements, the multimeter measures a DC reference voltage
applied to the W Sense terminals and a signal voltage applied to the Input
terminals. The multimeter then computes the ratio as:
The signal voltage measurement function can be DC voltage, AC voltage, or
AC+DC voltage. (For AC or AC+DC voltage, any of the three measurement
methods ANA, RNDM, or SYNC may be used.) The reference voltage
measurement function is always DC voltage and has a maximum measurable
input of ± 12VDC. Figure 15 shows the front connections for ratio
measurements
Note The W Sense LO and the Input LO terminals must have a common
reference and cannot have a voltage difference greater than 0.25V
.
a
tio
Signal Voltage
DC Reference Volta
g
------------------------------------------------
--
=
Figure 15. Ratio measurement connections

Table of Contents

Other manuals for Agilent Technologies 3458A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 3458A and is the answer not in the manual?

Agilent Technologies 3458A Specifications

General IconGeneral
BrandAgilent Technologies
Model3458A
CategoryMultimeter
LanguageEnglish

Related product manuals