Appendix A Specifications 291
Reading Rates
1
Settling Characteristics
For first reading or range change error using default delays, add .01% of input step additional error.
The following data applies for DELAY 0.
Maximum Input
Random Sampled Mode (ACV Function, SETACV RNDM)
AC Accuracy
3
24 Hour to 2 Year (% of Reading + % of Range)
Sec / Reading
ACBAND Low NPLC ACV ACDCV
³10 Hz 10 1.2 1
³1 kHz 11 0.1
³10 kHz 0.1 1 0.02
Function ACBAND Low DC Component Settling Time
ACV ³ 10 Hz DC < 10% AC 0.5 sec to 0.01%
DC > 10% AC 0.9 sec to 0.01%
ACDCV 10 Hz–1 kHz 0.5 sec to 0.01%
1 kHz–10 kHz 0.08 sec to 0.01%
³10 kHz 0.015 sec to 0.01%
Related Input Non-Destructive
HI to LO ±1000 V pk ±1200 V pk
LO to Guard ±200 V pk ± 350 V pk
Guard to Earth ± 500 V pk ±1000 V pk
Volt – Hz
Product
1 × 10
8
Range Full Scale Maximum
Resolution
Input Impedance
(Temperature Coefficients
2
% of Reading+% of Range)/°C
10 mV 12.000 1 µV
1 MW
±15% with<140 pF
0.002 + 0.02
100 mV 120.00 10 µV 1 MW ±15% with<140 pF 0.001 + 0.0001
1 V 1.2000 100 µV 1 MW ±15% with<140 pF 0.001 + 0.0001
10 V 12.000 1 mV 1 MW ±2% with<140 pF 0.001 + 0.0001
100 V 120.00 10 mV 1 MW ±2% with<140 pF 0.0015 + 0.0001
1000 V 700.0 100 mV 1 MW ±2% with<140 pF 0.001 + 0.0001
ACBAND £ 2 MHz ACBAND > 2 MHz
Range
20 Hz
to 100 kHz
100 kHz
to
300 kHz
300 kHz
to
1 MHz
1 MHz
to
2 MHz
20 Hz
to
100 kHz
100 kHz
to
1 MHz
1 MHz
to
4 MHz
4 MHz
to
8 MHz
8 MHz
to
10 MHz
10 mV 0.5+0.02 4+0.02 0.1+0.05 1.2+0.05 7 + 0.07 20 + 0.08
100 mV-10 V 0.08+0.002 0.3+0.01 1+0.01 1.5+0.01 0.1 +0.05 2+0.05 4 + 0.07 4 + 0.08 15 + 0.1
100 V 0.12+0.002 0.4+0.01 1.5+0.01 0.12+0.002
1000 V 0.3+0.01 0.3+0.01
1. For DELAY–1: ARANGE
OFF
For DELAY 0; NPLC .1 ,
unspecified reading rates of
greater than 500/Sec are
possible.
2. Additional error beyond±1°
C. but within ±5°C of last
ACAL.
For ACBAND > 2 MHz,
use 10 mV range
temperature coefficient for
all ranges.
3. Specifications apply from
full scale to 5% of full
scale. DC
< 10% of AC, sine wave
input, crest factor=1.4, and
PRESET. Within 24 hours
and ±1°C of last ACAL. LO
to Guard switch on.
Add 2 ppm of reading
additional error for Agilent
factory traceability of 10V
DC to US NIST.
Maximum DC is limited to
400V in ACV function.
Common Mode Rejection
For 1 kW imbalance in LO lead, > 90 dB, DC – 60 Hz.

Table of Contents

Other manuals for Agilent Technologies 3458A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 3458A and is the answer not in the manual?

Agilent Technologies 3458A Specifications

General IconGeneral
BrandAgilent Technologies
Model3458A
CategoryMultimeter
LanguageEnglish

Related product manuals