292 Appendix A Specifications
AC + DCV Accuracy (ACDCV Function)
For ACDCV Accuracy apply the following additional error to the ACV accuracy. (% of Range).
Additional Errors
Apply the following additional errors as appropriate to your particular measurement setup. (% of Reading)
Reading Rates
3
Settling Characteristics
For first reading or range change error using default
delays, add 0.01% of input step additional error.
The following data applies for DELAY 0.
DC £10% of AC Voltage DC >10% of AC Voltage
Range
ACBAND
£ 2MHz
ACBAND
>2 MHz
Temperatu re
Coefficient
1
ACBAND
£ 2MHz
ACBAND
>2 MHz
Temperature
Coefficient
1
10 mV 0.09 0.09 0.03 0.7 0.7 0.18
100 mV–1 kV 0.008 0.09 0.0025 0.07 0.7 0.025
Input Frequency
2
Source R 0–1 MHz 1–4 MHz 4–8 MHz 8–10 MHz
0 W 025 5
50 W Terminated 0.003 0 0 0
75 W Terminated 0.004 2 5 5
50 W 0.005 3 7 10
Sec/Reading
% Resolution ACV ACDCV
0.1 – 0.2 40 39
0.2 – 0.4 11 9.6
0.4 – 0.6 2.7 2.4
0.6 – 1 1.4 1.1
1 – 2 0.8 0.5
2 – 5 0.4 0.1
>5 0.32 0.022
Function DC Component Settling Time
ACV DC < 10% of AC 0.5 sec to 0.01%
DC > 10% of AC 0.9 sec to 0.01%
ACDCV No instrument settling required.
Crest Factor Resolution MultIplier
1–2 (Resolution in %) × 1
2–3 (Resolution in %) × 3
3–4 (Resolution in %) × 5
4–5 (Resolution in %) × 8
High Frequency Temperature Coefficient
For outside Tcal ±5°C add the following error.
(% of Reading) /
°C
Range 2– 4 MHz 4– 10 MHz
10 mV – 1 V 0.02 0.08
10 V – 1000 V 0.08 0.08
1. Additional error beyond ±1°C,
but within ±5°C of last
ACAL.
(% of Reading) / °C.
For ACBAND > 2 MHz, use
10 mV range temperature
coefficient for all ranges.
2. Flatness error including
instrument loading.
3. For DELAY –1;ARANGE
OFF. For DELAY 0 in ACV,
the reading rates are identical
to ACDCV.
Common Mode Rejection
For 1 kW imbalance in LO lead, > 90 dB, DC to 60
Hz.
Maximum Input
Rated Input Non-Destructive
HI to LO ±1000 V pk ±1200 V pk
LO to Guard ± 200 V pk ± 350 V pk
Guard to Earth ± 500 V pk ±1000 V pk
Volt – Hz
Product 1 x 10
8