TONE
256 Chapter 6 Command Reference
PRESET state, the multimeter uses the NRDGS command. The power-on values
for SWEEP can only be used for sub-sampling (since NRDGS does not apply
to sub-sampling).
• You cannot use the TIMER (or SWEEP) event for AC or AC+DC voltage
measurements using the synchronous or random methods (SETACV SYNC or
RNDM) or for frequency or period measurements.
• Query Command. The TIMER? query command returns the present time
interval, in seconds, for the NRDGS timer event.
• Related Commands: DELAY, NRDGS, SWEEP
Example 10 OUTPUT 722;"TRIG HOLD" !SUSPENDS MEASUREMENTS
20 OUTPUT 722;"INBUF ON" !ENABLES THE INPUT BUFFER
30 OUTPUT 722;"DCV 10" !DC VOLTAGE, 10V RANGE
40 OUTPUT 722;"NPLC .1" !SELECTS .1 PLC OF INTEGRATION TIME
50 OUTPUT 722;"AZERO OFF" !DISABLES AUTOZERO
60 OUTPUT 722;"MEM FIFO" !ENABLES READING MEMORY (FIFO MODE)
70 OUTPUT 722;"TIMER 2" !SELECTS 2 SECOND INTERVAL
80 OUTPUT 722;"NRDGS 10 TIMER" !10 READINGS PER SAMPLE EVENT (TIMER)
90 OUTPUT 722;"TRIG SGL" !TRIGGERS ONCE
100 END
TONE
Causes the multimeter to beep once. The multimeter then returns to the previous
BEEP mode (either OFF or ON).
Syntax TONE
Related Commands: BEEP
Example OUTPUT 722; "TONE" ! BEEPS
TRIG
Specifies the trigger event.
Syntax TRIG [event]
event