EasyManuals Logo
Home>EDAN>Monitor>iM50

EDAN iM50 User Manual

EDAN iM50
317 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #161 background imageLoading...
Page #161 background image
Patient Monitor User Manual Monitoring CO
2
- 149 -
3. Select Zero Calibration in CO
2
Setup menu.
4.
After the zeroing calibration is completed, the zeroing message disappears, and the CO
2
monitoring can be performed.
For the Respironics Sidestream CO
2
Module:
1.
Connect the sample line to the module correctly, wait until the monitor’s warm-up message
disappears, keep the inlet of sample line away from CO
2
source.
2.
In the CO
2
Setup menu, set Work Mode to Measure.
3.
Select Zero Calibration in CO
2
Setup menu.
4.
After the zeroing calibration is completed, the zeroing message disappears, and the CO
2
monitoring can be performed. If the monitor displays
Breath Detected or Zero Required,
zeroing has failed. Zero calibration must be performed again.
For the Masimo Sidestream CO
2
module:
The highly stable Masimo Sidestream CO
2
module requires no regular zeroing. A room air
reference measurement is performed when the NomoLine is disconnected from the gas inlet,
provided that CO
2
measurements are stable. This zeroing procedure is indicated by the blinking
green.
For the Respironics Mainstream CO
2
Module:
1.
Wait until the monitor’s warm-up message disappears; correctly install the mainstream CO
2
sensor to airway adaptor and remove it from breathing circuit, keep the monitor away from
CO
2
source.
2.
In the CO
2
Setup menu, set Work Mode to Measure.
3.
Select Zero Calibration in CO
2
Setup menu.
4.
After the zeroing calibration is completed, the zeroing message disappears, and the CO
2
monitoring can be performed. If the monitor displays
Breath Detected or Zero Required,
zeroing has failed. Zero calibration must be performed again.
Note: CO
2
source includes ventilator, patient’s and operator’s breath.
For Masimo Mainstream CO
2
Module:
The user can click
Zero button in CO
2
Setup, during the zeroing progress, the monitor displays
prompt information of “
Zeroing”.
Zeroing needs to be performed only when an offset in gas value is observed, or when an
unspecified accuracy message is displayed.
17.3.2 Sidestream CO
2
Module
17.3.2.1 Measurement Steps
For the EDAN EtCO
2
Module:
1. Fix the water trap to the water trap holder on the left side of the monitor, confirm it is well

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EDAN iM50 and is the answer not in the manual?

EDAN iM50 Specifications

General IconGeneral
BrandEDAN
ModeliM50
CategoryMonitor
LanguageEnglish

Related product manuals