Patient Monitor User Manual Default Settings
- 295 -
PVC On Medium Off
BIGEMINY On Medium Off
TRIGEMINY On Medium Off
TACHY On Medium Off
BRADY On Medium Off
MISSEDBEATS On Medium Off
IRR On Medium Off
PNC On Medium Off
PNP On Medium Off
VBRADY On Medium Off
VENT On Medium Off
C.4 RESP Default Settings
RESP Settings ADU PED NEO
Alarm Switch On
Alarm Record Off
Alarm Level Medium
Alarm High Limit 30 30 100
Alarm Low Limit 8 8 30
Apnea Alarm Time 20 s
Calculation Type Auto
Resp Type II
Sweep 12.5 mm/s
Amplitude 1
C.5 SpO
2
Default Settings
SpO
2
Settings ADU PED NEO
Alarm Switch On
Alarm Record Off
Alarm Level Medium
Alarm High Limit 100 100 95