Patient Monitor User Manual Default Settings
- 296 -
Alarm Low Limit 90 90 88
Pitch Tone On
Sensitivity Medium
SatSeconds (Nellcor Module) Off
Sweep 12.5 mm/s
SpO
2
Desat Limit 80%
C.6 PR Default Settings
PR Settings ADU PED NEO
PR Source SpO
2
Alarm Switch On
Alarm Record Off
Alarm Level Medium
Alarm High Limit 120 160 200
Alarm Low Limit 50 75 100
Pulse Volume 3
Alarm Source Auto
C.7 NIBP Default Settings
NIBP Settings ADU PED NEO
Alarm Switch On
Alarm Record Off
Alarm Level Medium
Alarm High Limit (SYS) 160 120 90
Alarm Low Limit (SYS) 90 70 40
Alarm High Limit (Map) 110 90 70
Alarm Low Limit (Map) 60 50 30
Alarm High Limit (Dia) 90 70 60
Alarm Low Limit (Dia) 50 40 20
Venipuncture pressure 60 40 30