EasyManuals Logo
Home>EDAN>Monitor>iM50

EDAN iM50 User Manual

EDAN iM50
317 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #82 background imageLoading...
Page #82 background image
Patient Monitor User Manual Alarm Information
- 70 -
Message Cause Alarm Level Action Taken
Network Disconnect
In distributed alarm system,
the monitor’s network is
disconnected.
Low
1) Check if the
network cable is well
connected.
2) Check if the
MFM-CMS is turned
on.
3) Check if the IP of
bedside monitor and
MFM-CMS are on
the same network
segment.
Network Traffic
Abnormity
Abnormal network traffic
has been detected. The data
traffic exceeds the limit.
High
Disconnect the
network to make the
monitor work
properly, and then
contact the
professionals
authorized by
manufacturer to
check the network
problem.
6.3 Prompts
Message Cause
ECG ARR Learning
The QRS template building required for Arr. Analysis is in
process.
Unable to analyze ST
The ST algorithm cannot produce valid ST value, which
may be caused by the large change in the measured value
of connected cardiogram ST or ventricular pacing.
Unable to analyze QT
QT algorithm cannot generate valid QT for more than 10
minutes (or 1 minute during startup).
QT Baseline Overrange
After modifying the calculation formula, the QTc
parameter value exceeds the range.
Unable to analyze ECG
The arrhythmia algorithm cannot analyze ECG data
reliably.
V-Fib/V-Tach (VFIB/VTAC)
Off
V-Fib/V-Tach (VFIB/VTAC) alarm is set to Off.
ExtremeTachy Off
Extreme Tachycardia alarm is set to Off.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EDAN iM50 and is the answer not in the manual?

EDAN iM50 Specifications

General IconGeneral
BrandEDAN
ModeliM50
CategoryMonitor
LanguageEnglish

Related product manuals