Patient Monitor User Manual Other Functions
- 208 -
Chapter 26 Other Functions
26.1 Nurse Call
The monitor provides dedicated nurse call port which is connected to nurse call system through
the nurse call cable to perform the nurse call function. You should activate the function following
the steps below:
1.
Select Menu > Maintenance > User Maintain, and input the password;
2.
Select Other Setups > Aux Output;
3. Choose On in the Nurse Call list.
NOTE:
Before using nurse call function, please check whether it is functioning normally.
26.2 Analog Output and Defibrillator Synchronization
The monitor provides analog output signals to accessory equipment. Also, if a defibrillator is
connected to the monitor, a defibrillator synchronization pulse can be output. You should activate
the function following the steps below:
1.
Select Menu > Maintenance > User Maintain, and input the password;
2.
Select Other Setups > Aux Output;
3. Choose Analog Output or Defibrillation in the Aux Output list.
26.3 Storing Data in the Storage Device
26.3.1 Data Stored in the Storage Device
Refer to Section Data Management for more information about single patient data volume.
You can choose to
Keep Storing or Stop Storing by selecting Menu > Common Function >
Data Store
> if one patient data full. When the single patient data reach the maximum, the
monitor will keep storing or stop storing as selected.
If you choose
Keep Storing, as soon as the single patient data is full, the earliest data will be
replaced by the latest one. When the remaining storage space is less than 15 M, the earliest
patient data in the storage space will be deleted in order to store the latest data.
If you choose
Stop Storing, the monitor will stop data storing and the latest data cannot be stored
when the single patient data reach the maximum. For instance, if all the patient data (such as the
trend graph, trend table, NIBP measurements, arrhythmia event, alarm event and 12-lead analysis)
except waveforms reach the maximum, the monitor will stop storing, while only the waveforms
keep storing until they are full. When the remaining storage space is less than 10 M, the monitor
will stop storing new data, prompting insufficient storage space. The default setting of
if one
patient data full
is Stop Storing.