Patient Monitor User Manual Networked Monitoring
- 38 -
is in the Wi-Fi Setup window, switch the Network Type from Wi-Fi to Wired and then
to Wi-Fi again. Then retry to connect to a wireless network. If the wireless network still
fails to be connected, please try to restart the monitor and connect again.
4 Use the wireless device recommended by EDAN, otherwise some exceptional
situations such as frequent network disconnection may occur on the monitor.
5 The wireless driver is compatible with channels 1-11 only.
6 When signal intensity is level 2 or less, signal may be unstable and quality of the
signal transmission may be degraded.
7 When the monitor is connected to MFM-CMS/Gateway via the wireless network, the
user should set the router to a secure encryption/authentication and use the
non-dictionary password.
Recommended options: WPA/WPA2 Personal (supports AES/TKIP);
Other options: none or WPA/WPA2 Enterprise (includes TLS/TTLS /PEAP).
8 The storage path for certificate related to EAP-TLS enterprise-level ecryption method:
root directory of USB flash drive\certs\wlan\.
WARNING
1 Before monitoring patient, Network Type (Wired or Wi-Fi) should be selected and is
not allowed to switch during monitoring. Otherwise, Wi-Fi may be unavailable.
2 If Wi-Fi is unavailable, please restart the monitor to restore Wi-Fi function under the
precondition of ensuring patient’s safety.
4.3 Network Disconnected Alarms
To configure the network disconnected alarms, select Menu > Maintenance > User Maintain >
Alarm Setup and choose Disconnect Alarm which can be set to On or Off. The alarm is off by
default.
NOTE:
1. When the monitor is connected with the central monitoring system, you must set
Disconnect Alarm to On.
2. If Disconnect Alarm occurs during audio alarm paused or audio alarm off status, the
monitor will prompt a sounding alarm with information of NetWork Disconnect.
During the network disconnected status, activating audio alarm paused or audio alarm
off function can disable the audio alarm signal of Disconnect Alarm.
4.4 Connecting the Monitor to MFM-CMS
The monitor can be connected to the central monitoring system (V2.65 and above). Through the
network:
1. The monitor sends patient information, real-time monitoring or measurement data to the
central monitoring system.