Patient Monitor User Manual Alarms
- 45 -
When the alarm pause time expires, the audio alarm paused status is automatically terminated and
alarm is sounding. You can also terminate the alarm paused status by pressing the hardkey
or pressing shortcut key .
NOTE:
If a new alarm occurs during the audio alarm paused period, the new alarm will not be
sounding.
5.4.3 Audio Alarm off
Set Pause Time to Permanent, press hardkey or shortcut key , the monitor
displays information: please confirm whether to activate audio alarm off function? Click Yes,
the monitor will enter into audio alarm off status. Click No, the monitor will keep current status.
• The audio alarm is turned off, and no alarms are sounding.
• The visual alarm indications are still displayed.
• The hardkey
on the front panel flashes in yellow.
Remind signal: Audio alarm off symbol and Audio Alarm off on a red colored
background are displayed with an interval of 2 s during the audio alarm off status.
Pressing the hardkey or shortcut key again can resume the audio alarm.
NOTE:
If a new alarm occurs during the audio alarm off period, the new alarm will not be
sounding.
5.4.4 Alarm Reset
Select the shortcut key on the screen directly. When the alarm is reset,
No alarms are sounding until a new alarm occurs.
As for the active alarms, the visual alarm indications are still displayed.
All latching alarms are cleared. If the alarm condition is no longer present, all alarm
indications stop and the alarm is reset.
It will not influence the configuration of physiological alarm off, audio paused, and audio off
status.
NOTE:
If a new alarm occurs after the alarm is reset, the new alarm will be sounding.
5.5 Latching Alarms
To configure the alarm latching setting, select Menu > Maintenance > User Maintain > Alarm