Patient Monitor User Manual Using Battery
- 214 -
3. Disconnect monitor from mains power and let the monitor run until there is no battery power
left and the monitor shuts off.
4.
The running time of the battery reflects the battery performance.
If the running time is obviously less than the specified time in the specification, please change the
battery or contact the service personnel. If the running time meets the specification, fully charge
the battery again for use or charge it to 40%-60% for storage.
27.5 Replacing the Battery
To install or replace the battery, please follow the procedure:
1.
Pull the battery door according to indication on it to open it.
2.
Pull the plastic retainer until the battery can be removed.
3.
Insert the new battery into the battery compartment.
4.
Pull the metal retainer downward to fix the battery and close the battery door.
27.6 Recycling the Battery
When the battery no longer holds a charge, it should be replaced. Remove the old battery from
the monitor and recycle it properly.
WARNING
Do not disassemble batteries, put them into fire or cause them to short circuit. They may
ignite, explode or leak, causing personal injury.
27.7 Maintaining the Battery
Batteries should be conditioned regularly to maintain their useful life.
Remove the batteries from the monitor if they are not used for a longer period of time.
It is recommended to check and maintain the batteries regularly every 3-6 months if they are not
Battery Door