Recommended module position
ConstraintsRecommended locationModule
Column valve V9-C or V9H-C cannot be
used at the same time as Column valve
V9-Cs or V9H-Cs.
9Column valve V9-C or
V9H-C
Column valve V9-Cs or V9H-Cs cannot
be used at the same time as Column
valve V9-C or V9H-C.
9Column valve V9-Cs or
V9H-Cs
Conductivity monitor C9 has no con-
straints.
7Conductivity monitor C9
Core module.6Injection valve V9-Inj or
V9H-Inj
Possible combinations of inlet valves:
•
4
•
5, if Inlet valve V9-IAB
or V9H-IAB is used in
combination with Inlet
valve V9-IA or V9H-IA.
Inlet valve V9-IAB or V9H-
IAB
•
Inlet valve A and Inlet valve B.
•
Inlet valve AB together with Inlet valve
A or Inlet valve B.
•
One Inlet valve AB.
or
•
No installed inlet valves.
4Inlet valve V9-IA or V9H-IA
5Inlet valve V9-IB or V9H-IB
Cannot be used at the same time as Mixer
valve V9-M or V9H-M.
8Sample inlet valve V9-IS or
V9H-IS
Inlet valve V9-X1 or V9H-X1 has no con-
straints, and has no support in the
method phases.
NoneInlet valve V9-X1 or V9H-
X1
Inlet valve V9-X2 or V9H-X2 requires Inlet
valve V9-X1 or V9H-X1, respectively. It has
no support in the method phases.
NoneInlet valve V9-X2 or V9H-
X2
Loop valve V9-L or V9H-L has no con-
straints, but is connected with the injec-
tion valve by replacing the loop.
8Loop valve V9-L or V9H-L
Core module.0Mixer M9
Cannot be used at the same time as
Sample pump S9 or S9H.
3Mixer valve V9-M or V9H-
M
Cannot be used at the same time as
Outlet valve V9-O or V9H-O.
1Outlet valve V9-Os or V9H-
Os
Cannot be used at the same time as
Outlet valve V9-Os or V9H-Os.
1Outlet valve V9-O or
V9H-O