4.2.7 Configuration of outlet valves
System properties
Follow the instruction below to update the system properties.
ActionStep
Open the system properties Edit dialog.1
Select Valves and pumps from the Component types list.2
Select components and properties according to the following table:3
PropertyComponent selectionValve
V9-O (10-outlets)Outlet valveOutlet fractionation valve
V9-O
V9H-O (10-outlets)Outlet valveOutlet fractionation valve
V9H-O
V9-Os (1-outlet)Outlet valveOutlet control valve V9-Os
V9H-Os (1-outlet)Outlet valveOutlet control valve V9H-
Os
System settings
If the valve is not placed in the recommended position using the standard tubing kit, the delay volume
must be set. See Check/Set delay volume, on page 146.
ÄKTA pure User Manual 29119969 AB 139
4 System configuration
4.2 Configure modules
4.2.7 Configuration of outlet valves