2.4.2 Mixer
Function of the Mixer
Mixer M9 is located after System pump A and System pump B, and before the Injection valve. The
Mixer is a dynamic mixer for high-performance gradients. It is used to make sure that the buffers from
the System pumps are mixed to give a homogenous buffer composition.
The Mixer has a built-in filter that prevents particles from reaching the column.
Replace the sealing ring of the mixer with a highly resistant O-ring if the system is going to be
exposed to organic solvents or high concentrations of organic acids for longer periods of time.
Refer to Section 7.8.4 Replace the O-ring inside the Mixer, on page 323 for how to change the
ring.
Note:
Location and illustration
The illustration below shows the location, together with a detailed view of Mixer M9.
DescriptionPart
Outlet1
Mixer chamber (0.6, 1.4, 5 or 15 ml)2
CAUTION
Risk of explosion. Do not use Mixer chamber 15 ml with
an ÄKTA pure 25 configuration. The maximum pressure
for Mixer chamber 15 ml is 5 MPa.
Inlet3
38 ÄKTA pure User Manual 29119969 AB
2 The ÄKTA pure instrument
2.4 Instrument modules
2.4.2 Mixer