2.4.5 Mixer valve
Function of Mixer valve
Mixer valve (V9-M or V9H-M) allows the user to bypass the mixer. It is intended to be used when the
System pump is used for sample application or when a sample is re-injected.
Mixer valve (V9-M or V9H-M) cannot be used together with Sample pump S9 or Sample pump
S9H.
Note:
Location and illustration of Mixer valve
The illustration below shows the recommended location, together with a detailed view of the Mixer
valve.
Ports and flow paths of the mixer valve
The illustration and tables below describe the ports of and different flow paths through the Mixer valve.
The Mixer valve has two available flow paths; By-pass, and Mixer (default). If the Mixer valve is installed
in the recommended location before the Mixer, By-pass allows the flow to bypass the Mixer, and Mixer
directs the flow to the Mixer.
ÄKTA pure User Manual 29119969 AB 47
2 The ÄKTA pure instrument
2.4 Instrument modules
2.4.5 Mixer valve