5.6 Pressure alarms
Introduction
The columns can be protected by two different types of pressure alarms:
The pre-column pressure alarm protects the column hardware
The delta-column pressure alarm (only available when V9-C or V9H-C is installed) protects the column
media
Column valves V9-C and V9H-C have built-in pressure sensors that automatically measure the pre-
column and delta-column pressure. If Column valve V9-C or V9H-C is not used (column is connected
without a Column valve or to Column valve V9-Cs or V9H-Cs), the pre-column pressure is calculated
from the system pressure and tubing dimensions.
See the instructions below to set the pressure alarm for the column to be used in the run and, if appli-
cable, to set the parameters for the tubing dimensions.
Set tubing dimension parameters
to calculate pre-column pressure
For instruments where there is no pre-column pressure sensor, i.e. the column is connected without
a Column valve or to Column valve V9-Cs or V9H-Cs, the pre-column pressure is calculated from the
system pressure and tubing dimensions. Follow the instructions below to set the tubing dimension
parameters.
ActionStep
Select System:Settings in the System Control module.1
Result: The System Settings dialog opens.
Select Tubing and Delay Volumes
and
select Tubing: Injection valve to column
2
ÄKTA pure User Manual 29119969 AB 177
5 Operation
5.6 Pressure alarms