7.8.8 Replace the inlet filters
Maintenance interval
Replace the inlet filter when required, for example when the filters are clogged.
Required material
The following material is required:
Inlet filter set
Instruction
Follow the instruction to replace an inlet filter and a support net.
ActionStep
Pull off the inlet filter and the support net from the inlet filter holder.1
Fit the new support net and inlet filter, and press the filter into position.2
ÄKTA pure User Manual 29119969 AB 331
7 Maintenance
7.8 Replacement procedures
7.8.8 Replace the inlet filters